Features
1. Adopting the principle of photoresist (shading) method, high-precision laser sensors are used, with small size, high accuracy, and stable performance.
2. Suitable for offline and online detection on site, it can monitor the pollution level of particles in the oil system in real time.
3. Equipped with a built-in data analysis system, it can display the real data of particle size in each channel and automatically determine the sample grade.
4. The internal pipelines are made of stainless steel and PTFE materials.
5. Built in volume flushing and quick flushing modes are convenient for users to use the equipment and clean and maintain the instrument, with built-in ISO4406 NAS1638、SAE4095、GJB420A、GJB420B、GB/T14039、 OCT17216 and other particle pollution level standards.
6. Built in calibration function, can calibrate according to standards such as GB/T21540, ISO4402, GB/T18854, etc., and can calibrate the instrument by oneself.
7. 10.1-inch color touch screen. Input in both Chinese and English. Easy to operate.
8. Built in compact ceramic metering pump. Accurate sampling and measurement
9. Large storage, can be stored inside the instrument or on a USB drive.
10. You can view memory data, print historical data, and export to a USB drive.
11. Optional RS485 interface, which can be connected to computers or other devices for data monitoring and debugging.
12. Built in polymer battery, no need to connect to a power source in the field.
13. Optional built-in moisture saturation, PPM value, and temperature.
14. Third party calibration certificates and verification certificates can be issued through optional options.
15. Dual operating systems in both Chinese and English. Can achieve one click switching
16. Stainless steel panel, engineering plastic waterproof packaging box, suitable for complex working environments.
General parameters
·Light source: semiconductor laser
·Flow rate range: 0-150mL/min
·Sensitivity: 1 μ m or 4 μ m (c)
·Test sample viscosity: ≤ 350cSt
·Online pressure detection: 0.1-0.5Mpa (with optional pressure reducing device, the maximum pressure can reach 40Mpa)
·Particle size range: 1-500 μ m (using different types of sensors)
·Data Storage: Provides 1000 sets of data and USB storage space
·Limit coincidence error: 10000-40000 particles/ml
·Counting volume: 1-100ml
·Counting accuracy: ± 0.3 pollution levels
·Protection level: IP56
·Temperature collection T (optional): 1. Collection range: 1-100 degrees Celsius; 2. Measurement accuracy: 1 degree Celsius; 3. Response time: 30s
·Water activity collection aw (optional): 1 Collection range: 1-100% RH; 2. Measurement accuracy: 1% RH; 3. Response time: 30s
·Absolute moisture content AH (optional): 1. Collection range: 1-350ppm; 2. Measurement accuracy: 1ppm; 3. Response time: 30s
·Test interval: 1-24 hours
·Test sample temperature: 0-80 ° C
·Working temperature: -20-60 ° C
·Power supply: AC 220V ± 10%, 50/60Hz or DC12-40V
·Weight: 6.0kg
·Volume: 435x343x150mm