Light source: semiconductor laser
Part Number : LWT-5 |
Unit Price : USD [Please inquire] |
Lead time : in stock |
Inventory quantity : [Please inquire] |
Stock NO. : |
Email Us Request for Quotation |
LWT-5 oil particle counter is the upgraded product of LWT desktop series, the instrument is based on GB/T 18854-2002 (ISO11171-1999) and other standards. Adopting the counting principle of photo-resistance (shading) method, it is fully compliant with the corresponding national military standards, national standards and international standards. It can provide fast, accurate, reliable and repeatable test results and complete pollution monitoring and analysis report, which is suitable for the detection of particle contamination of hydraulic oil, lubricating oil, anti-fuel oil, insulating oil and turbine oil. It can be widely used in aviation, aerospace, electric power, petroleum, chemical industry, transport, harbour, metallurgy, machinery, automobile manufacturing and other fields.
- Light source: Semiconductor laser
- Particle size range: 0.8μm~600μm (calibrated according to ACFTD calibration 1~100μm or ISO MTD calibration 4~70μm(c))
- Detection channels: 8~64 channels optional (factory default 64), any set particle size size
- Sensitivity: 0.8μm (ISO4402) or 3μm (c) (GB/T18854, ISO11171)
- Resolution:<10% (GB/T18854, ISO11171)
- Repeatability: RSD<2%
- Sample viscosity: ≤650cSt (large viscosity can be detected with heating or dilution method)
- Sampling volume: 0.2~6000mL, interval 0.1mL
- Sampling accuracy: better than ± 0.5%
- Sampling speed: 5 ~ 80mL/min
- Barometric chamber: Positive and negative barometric chamber devices, to achieve sample degassing and high viscosity sample detection
- Maximum vacuum of barometric chamber: -0.08Mpa
- Maximum positive pressure of barometric chamber: 0.8 Mpa
- Maximum particle concentration: 12000~40000 particles/mL
- Temperature of test sample: 0℃~80
- Operating Temperature: -20℃~60℃
- Storage Temperature: -30℃~80℃
- Power supply: 110~245V AC, 50/60Hz, 70W
- Main unit size: 340mm×410mm×650mm Air pump size: 180mm×160mm×220mm
- Instrument memory: 8G